FP92 5G2 - Cleveland Open Cup Flash Point

Test Methods

ASTM D92, ISO 2592, IP 36, EN 22592, NF EN 22592, GOST 4333

 

Description

ISL’s FP92 5G2 combines leading flash point technology with user friendly automation in a conveniently compact design, which boosts your laboratory’s productivity while greatly improving repeatability and reproducibility. Operation couldn’t be easier; simply select from a list of pre-programmed testing conditions, then press the Test key. Test progress displays in real time on the easily viewable screen.

Sophisticated quality control featuresincluding traceable automatic calibrations with lock-out control and automated quality verification fulfill your ISO 9000 quality system requirements. Its extensive local data storage is further supplemented by LIMS export capabilities, while the addition of ALAN enables multi-instrument networking with centralized operation control and data management. For enhanced automation, add ISL’s SC6 6-position sample changer at any time. On-screen prompts guide you through the quick, simple test setup and allow you to program new sample positions even as another test is in progress. Each of the six test cups is equipped with its own heating plate; once a completed sample’s flame has been extinguished, the next test begins immediately with a cold plate.

 

Applications

Flash Point Cleveland Open Cup

 

Features & Benefits

Unparalleled Versatility, Absolute Simplicity

Easy test initiation by factory pre-installed test method; able to customize test methods and conditions

Real-time display during testing, which is viewable from 5 meters (16 ft.)

Multi-language user support

Automatic lighting and test flame monitoring

Optional cooling block for sub-ambient testing

Detachable cup cover and shutter simplify cleaning

Enhanced Quality Features

Built-in sensor for automatic barometric pressure correction according to ASTM or ISO methods

Automatic, time-recorded calibrations with lock-out control

Warning message when result is out of specification

Safe, Dependable Operation

Relights flame, if necessary, during operation and suppresses gas source at end of test

Automatic fire detection system with external fire alarm connection

Over-heating detection with automatic heat shut-off during test

Optional PC Control & Multi-Instrument Networking

Centralizes control and data management of up to 31 ISL ALAN-ready cold behavior, distillation & flash point instruments

Automatically stores results to database with sorting and filtering capabilities

Transmits results to external computer, network or LIMS with data extraction mask for each test program

© 2017 Sigmatech Inc.. All rights reserved.