Electron Beam Absorbed Current (EBAC) technique offers a quick and effective method to identify open circuits, high resistance and shorts along interconnects without direct probing techniques of lower level layers.
The EBAC technique is performed with the electron beam by passing through the dielectric layers down to the lower level metallization layer in order to absorb the electron beam current. The electron beam accelerating voltage of the FESEM controls the probing depth or penetration level through the dielectric layers. A single probe is placed on the exposed, upper layer metallization to complete the circuit and allow the electrons to flow through the interconnect.
Observation of high resistance and shorts due to Seebeck effect is possible by using dual probes along with the Hitachi patented differential EBAC amplifiers.
User-Friendly Design
- Intuitive GUI(Graphical User Interface) with various image and color processing functions.
- Coarse positioning of probes are accomplished by an integrated in-chamber CCD camera system.
Premium Image Quality
- Provides high quality EBAC images with Hitachi’s patented high performance EBAC amplifiers.
Outstanding Performance
- Field Proven, low chromatic aberration, Cold Field Emission (CFE) electron gun for low accelerating voltage imaging and beam damage reduction of the circuit.
- High precision nano-probe units.
Advanced Applications
- Dedicated nano-probing system with EBAC analysis and electrical characteristics analysis. Large specimen stage can provide fine material and electronics components evaluation.
Specification
Hitachi Electron Beam Absorbed Current(EBAC) Characterization System nanoEBAC NE4000
Probe unit |
4 |
Piezoelectric |
5 µm (X,Y) |
6 mm (X,Y) |
Specimen stage / Base stage |
25 mm × 25 mm × 1 mm thick or less |
Measurement / Specimen exchange position |
Air-locked exchange chamber |
Stage traverse to probe position |
Measurement position memory |
Probe coarse adjustment |
Image display from lateral direction |
Electron optics |
Cold field emission electron source |
0.5 kV to 30 kV |
15 nm (at 2 kV, WD=15 mm) |
±150 µm (at 2 kV, WD=15 mm) |
EBAC amplifier / Image display |
Current amplifier / Differential amplifier |
SEM / EBAC (Single / Parallel / Overlay) |
Black and white reversal display, color display, brightness adjustment, slow scan integration, belt scan |
Dimensions and Weight
Main unit |
1,100 (W) × 1,550 (D) × 1,750 (H) mm,
850 kg |
Display unit |
1,000 (W) × 1,005 (D) × 1,200 (H) mm,
265 kg |
Utility requirement
Room temperature |
15 – 25 °C |
Humidity |
60% RH or less |
Power |
AC100 V±10% 5 kVA (M5 crimp terminal) |
Grounding |
100Ω or less |