Data Logger

4-Channel Analog Data Logger

A high-performance, LCD display data logger to measure and record temperature, current, CO2, voltage, and more, for building performance monitoring applications.

4 Channel Analog Data Logger

Overview

This logger is ideal for energy audits, building commissioning studies, equipment scheduling, and more.

4 Channel Analog Data Logger

Features
  • Twice the accuracy of previous models with 16-bit resolution
  • Flexible support for a wide range of external sensors
  • LCD confirms logger operation and displays near real-time measurement data
  • Provides minimum, maximum, average, and standard deviation logging options
  • On-screen alarms notify you when a sensor reading exceeds set thresholds
  • Stores 1.9 million measurements for longer deployments between offloads
  • Compatible with HOBOware and HOBOware Pro software for logger setup and graphing & analysis

4 Channel Analog Data Logger

Specifications

Logger w/Cable

Measurement Range:
4-20mA (w/CABLE-4-20MA) 0 to 20.1 mA
0 to 2.5 V (w/CABLE-2.5-STEREO)
0 to 5 V (w/CABLE-ADAP5)
0 to 10 V (w/ CABLE-ADAP10)
0 to 24 V (w/ CABLE-ADAP24)

Accuracy:
4-20mA ±0.001 mA ±0.2% of reading (w/CABLE-4-20MA)
±0.1 mV ±0.1% of reading (w/CABLE-2.5-STEREO)
±0.2 mV ±0.3% of reading (w/CABLE-ADAP5)
±0.4 mV ±0.3% of reading (w/ CABLE-ADAP10)
±1.0 mV ±0.3% of reading(w/ CABLE-ADAP24)

Resolution:
0.3 µA 4-20mA (w/CABLE-4-20MA)
40 µV (w/CABLE-2.5-STEREO)
80 µV (w/CABLE-ADAP5)
160 µV (w/ CABLE-ADAP10)
384 µV (w/ CABLE-ADAP24)

4 Channel Analog Data Logger

Logger

Operating range logging: -20° to 70°C (-4° to 158°F); 0 to 95% RH (non-condensing);
Launch/readout: 0° to 50°C (32° to 122°F) per USB specification
Logging rate: 1 second to 18 hours, 12 minutes, 15 seconds
Logging modes: Normal, burst, or statistics
Memory modes: Wrap when full or stop when full
Start modes: Immediate, push button, date & time, or next interval
Stop modes: When memory full, push button, or date & time
Restart mode: Push button
Time accuracy: ±1 minute per month at 25°C (77°F), see Plot A
Battery life: 1 year, typical with logging rate of 1 minute and sampling interval of 15 seconds or greater
Battery type: Two AAA 1.5 V alkaline batteries, user replaceable
Memory: 4 MB (1.9 million measurements, maximum)
Download type: USB 2.0 interface
Full memory download time: approximately 1.5 minutes
LCD: LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range
Size: 10.8 x 5.41 x 2.54 cm (4.25 x 2.13 x 1 in.)
Weight: 107.5 g (3.79 oz)
Environmental rating: IP50

4 Channel Analog Data Logger

4 Channel Analog Data Logger

For more informationcontact us.

Choose user manual language:

 

Choose quick start language:

Features

Choose user manual language:

 

Choose quick start language:

Related Products

The LI-7700 is designed specifically for eddy covariance flux measurements to evaluate methane emissions from the terrestrial landscape.
The LI-1500 allows for simple configuration of automated sampling to compute daily light integral (DLI) from PAR readings captured by LI-COR quantum sensors.
The LAI‑2200C Plant Canopy Analyzer uses a non-destructive method to easily and accurately measure Leaf Area Index (LAI).
Measures and transmits temperature, relative humidity, and light intensity data wirelessly.
A battery-powered rainfall data collection and recording system that includes a HOBO Pendant Event data logger integrated into a tipping-bucket rain gauge.
A weatherproof data logger with an external soil/water temperature sensor on a 6-foot cable, for applications where just one temperature probe is required. Also includes an internal temperature sensor
Monitors room occupancy and indoor light changes to identify occupancy patterns and determine energy usage and potential savings.
Monitors indoor light changes with an internal sensor and records light on/off conditions; ideal for building energy audits and light usage analysis.
Rugged and Bluetooth-enabled, the MX2308 delivers research-grade monitoring of key growing conditions—photosynthetically active radiation (PAR), temperature and relative humidity—all in a single devic
This 15-channel data logger lets energy and facility managers monitor energy and industrial equipment to quickly and easily detect and address a variety of issues relating to HVAC, machinery

How can we help?

You may contact our specialists by accomplishing form below.

failure analysis, material characterization, and metrology equipment

Subscribe to newsletter

Sign up for our newsletter to get updates on promos, seminars, events, products, application notes and more.

Inquire for this Product

4-Channel Analog Data Logger
Data Logger
4-Channel Analog Data Logger
Data Logger