Performance
Excellent short and long term stability are found in our QuickTrace® M-7600 CVAA analyzer. Less than 0.3 ng/L instrument detection limits are typical for this analyzer while utilizing less than 4 minutes of sample time and less than 25 mL of sample.
- Ultra-trace detection limits (< 0.5 ng/L IDL)
- Linearity greater than 4 orders of magnitude
- Dynamic range < 0.5 ng/L to > 500 μg/L
- Without employing our smart rinse technology the system is ready to accurately measure a sample within four minutes following the ingestion of a 1 mg/L sample
- Short term precision (%RSD @ 95% Confidence) < 1.2% @ 20 ng/L, n=5
- Short term precision (%RSD @ 95% Confidence) < 6.0% @ 5 ng/L, n=5
Common Methods
- EPA Method 245.1
- EPA Method 245.5
- SW 846-7470
- SW 846-7471
- ASTM Method; D3223, D3624, D3684, D6414, and D6784
- EU Methods; Din ISO 16772, EN-13806, EN-16175-1, and ISO 12846